Simple Plex Human Amphiregulin Cartridge

Catalog #: SPCKB-PS-001181 Datasheet
Simple Plex Human Amphiregulin assay kit for use on Ella instrument. Contains cartridge, sample diluent SD13, and wash buffer
Catalog # Availability Size / Price Qty
SPCKB-PS-001181
SPCKA-PS-001181
SPCKC-PS-001181
Control Products Available
Human Amphiregulin Simple Plex Assay
3 Images
Product Details
FAQs
Supplemental Products
Reviews

Simple Plex Human Amphiregulin Cartridge Summary

Assay Type
Automated sandwich ELISA
Format
Microfluidic cartridge containing three Glass Nano Reactors (GNR)
Assay Length
90 minutes
Sample Type & Volume Required Per Well*
Serum (25 uL), EDTA Plasma (25 uL), Heparin Plasma (25 uL)
Sensitivity
2.62 pg/mL
Assay Range
5.4 - 51,610 pg/mL (Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant human Amphiregulin.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.

*The Sample Volume represented is based on the amount of sample incorporated into the reaction after taking into account the assay’s minimum required dilution for a given matrix. Serial dilution may be necessary to achieve some of the final sample volumes represented.

Product Summary

Simple Plex assay for the detection of human Amphiregulin in serum and plasma (EDTA/Heparin)

Precision

Intra-Assay Precision (Precision within an assay) Each control was tested 16 times in one assay
Inter-Assay Precision (Precision between assays) Replicates of each control were tested in multiple assays performed by at least three technicians using two lots of reagents

Serum, EDTA Plasma, Heparin Plasma

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 1 2
n 16 16 69 69
Mean (pg/mL) 73.7 3225 68.1 3228
Standard Deviation 3.19 236 7.01 285
CV% 4.3 7.3 10.3 8.8

Recovery

Recovery at three different spiked concentrations within the range of the assay was evaluated.

Sample Type Average % Recovery Range %
EDTA Plasma (n=4) 83 81-85
Heparin Plasma (n=4) 83 79-86
Serum (n=4) 84 83-85

Linearity

Samples containing and/or spiked with high concentrations of human Amphiregulin were serially diluted with Sample Diluent to produce samples within the dynamic range of the assay
Human Amphiregulin Simple Plex Assay Linearity

Scientific Data

Human Amphiregulin Simple Plex Assay Standard Curve

Product Datasheets

You must select a language.

x

Preparation and Storage

Shipping
The product is shipped at ambient temperature. Upon receipt, store it immediately at the temperature recommended below.
Storage
Store the unopened product at 2 - 8 °C. Do not use past expiration date.

Background: Amphiregulin

Amphiregulin (AR) is an EGF family growth factor that is released as a soluble protein following proteolytic cleavage of its transmembrane precursor. Amphiregulin is expressed by numerous carcinoma cell lines and epithelial cells of the colon, stomach, breast, ovary and kidney. It acts through ErbB family of receptor to stimulate the proliferation of keratinocytes, mammary epithelial cells, fibroblasts, astrocytes and glial cells. Amphiregulin can also inhibit the growth of certain tumor cells.

Entrez Gene IDs:
374 (Human); 11839 (Mouse)
Alternate Names:
Amphiregulin; AR; AREG; AREGB; Colorectum cell-derived growth factor; CRDGF; MGC13647; schwannoma-derived growth factor; SDGF

FAQs

No product specific FAQs exist for this product, however you may

View all Simple Plex Assays FAQs
Loading...

Reviews for Simple Plex Human Amphiregulin Cartridge

There are currently no reviews for this product. Be the first to review Simple Plex Human Amphiregulin Cartridge and earn rewards!

Have you used Simple Plex Human Amphiregulin Cartridge?

Submit a review and receive an Amazon gift card.

$25/€18/£15/$25CAN/¥75 Yuan/¥2500 Yen for a review with an image

$10/€7/£6/$10 CAD/¥70 Yuan/¥1110 Yen for a review without an image

Submit a Review