Simple Plex Discovery Human TRANCE/RANK L/TNFSF11 Assay New
Simple Plex Discovery Human TRANCE/RANK L/TNFSF11 Assay Summary
*The Sample Volume represented is based on the amount of sample incorporated into the reaction after taking into account the assay’s minimum required dilution for a given matrix. Serial dilution may be necessary to achieve some of the final sample volumes represented.
Product Summary
Precision
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma
Intra-Assay Precision | Inter-Assay Precision | |||
---|---|---|---|---|
Sample | 1 | 2 | 1 | 2 |
n | 16 | 16 | 24 | 24 |
Mean (pg/mL) | 198 | 9796 | 253 | 12325 |
Standard Deviation | 8.65 | 482 | 28.1 | 1286 |
CV% | 4.4 | 4.9 | 11.1 | 10.4 |
Recovery
Recovery for the human TRANCE assays was evaluated at three different spiked concentrations of TRANCE protein.
Sample Type | Average % Recovery | Range % |
---|---|---|
Cell Culture Media (n=2) | 112 | 107-117 |
EDTA Plasma (n=2) | 98 | 81-116 |
Heparin Plasma (n=2) | 114 | 109-120 |
Serum (n=2) | 108 | 101-115 |
Linearity
Scientific Data

Data shown represents typical performance results for Lower Limit of Quantitation (LLOQ) and Upper Limit of Quantitation (ULOQ) for Simple Plex Human TRANCE Assay. Assay range is 21.0-32,000 pg/mL.
Product Datasheets
Preparation and Storage
Background: TRANCE/TNFSF11/RANK L
TRANCE (TNF-related activation-induced cytokines), also known as RANK Ligand, OPGL, or ODF, is a TNF superfamily ligand produced by osteoblasts, bone marrow stromal cells, and activated T cells. It signals through RANK and is antagonized by the decoy receptor Osteoprotegerin. TRANCE stimulates bone resorption, is required for osteoclast differentiation, and is associated with the development of osteoporosis, vascular calcification, osteopetrosis, and progestin-induced breast cancer. It also mediates body temperature regulation and is essential for the development of a lactating mammary gland.
FAQs
No product specific FAQs exist for this product, however you may
View all Simple Plex Assays FAQsReviews for Simple Plex Discovery Human TRANCE/RANK L/TNFSF11 Assay
There are currently no reviews for this product. Be the first to review Simple Plex Discovery Human TRANCE/RANK L/TNFSF11 Assay and earn rewards!
Have you used Simple Plex Discovery Human TRANCE/RANK L/TNFSF11 Assay?
Submit a review and receive an Amazon gift card.
$25/€18/£15/$25CAN/¥75 Yuan/¥2500 Yen for a review with an image