Human CD14 Magnetic Luminex® Performance Assay
Build your own Luminex Assay with our Luminex Assay Customization Tool.
U.S. orders usually ship in 2-3 business days
Build NowHuman CD14 Magnetic Luminex® Performance Assay Summary
Panel | Luminex Performance Human Biomarker Panel A |
Assay Type | Magnetic bead-based multiplex assay for the Luminex® platform |
Format | 1 x 96-well microplate and Magnetic antibody-coated beads |
Species | Human |
Analytes Detected | Please see analyte list in assay customization tool below. |
Performance Validation | Our Luminex High Performance Assays undergo our most extensive validation testing. All Luminex High Performance Assay Panels are validated for use with serum and plasma samples. Some High Performance Assays are additionally validated for cell culture supernatants, milk, saliva, or urine, as indicated on the individual product data sheets. Luminex High Performance Assays are tested for sensitivity (three-quarters the low standard), intra-assay precision, inter-assay precision, and to ensure assay linearity for validated sample types. Recovery values for individual samples in validated sample types are also tested. Validation data for each analyte can be found in the product datasheet. |
Make your selections below to configure your assay
Not finding your analyte of interest? View our complete Luminex offering in the full version of our Luminex Assay Ordering Tool. We also offer Custom Luminex Services
Magnetic Luminex High Performance Assays are flexible bead-based multiplex assays. They allow up to 50 user-defined target analytes to be simultaneously profiled using cell culture supernates, serum, or plasma samples. Some panels are also validated with human milk, saliva, and/or urine samples.
- Simultaneously profile Human CD14 with up to analytes of your choice in one sample
- Magnetic format allows for easier wash steps
- Undergoes similar validation testing as Quantikine® ELISA Kits
- Precise and Specific
- Requires a small sample volume (<50 µL)
- Run your assay in just 3-5 hours
- Mass-calibrated standards for consistent results with every new lot of material
- Enough reagents for one 96-well plate
- Standard Cocktail(s)
- Bead Diluent
- Biotin Antibody Diluent
- Calibrator Diluent
- Wash Buffer
- Streptavidin-PE
- One flat-bottom 96-well Microplate
- Foil Plate Sealers (4)
- Mixing Bottle
- Standard Value Card**
**A standard curve must be generated each time an assay is run, utilizing values from the Standard Value Card included in the Base Kit.
- User-mix antibody-coated Magnetic beads in individual vials
- User-mixed cocktail of biotinylated detection antibody
Precision
Intra-assay Precision (precision within an assay) - Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision. |
||||||
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 32 | 32 | 32 |
Mean (pg/mL) | 593 | 4427 | 25,835 | 591 | 4756 | 32,468 |
Standard deviation | 33.2 | 172 | 2489 | 85.3 | 540 | 4287 |
CV (%) | 5.6 | 3.9 | 9.6 | 14.4 | 11.3 | 13.2 |
Recovery
Samples containing and/or spiked with human CD14 were evaluated for recovery. |
||||||
Sample Type | Average % Recovery | Range | ||||
---|---|---|---|---|---|---|
Cell culture supernates | 100 | 93-114% |
Linearity
Samples containing and/or spiked with human CD14 were serially diluted to evaluate assay linearity. |
|||||||||
Cell culture supernates* | Serum | EDTA plasma | Heparin plasma | ||||||
---|---|---|---|---|---|---|---|---|---|
1:2 | Average % of Expected | 106 | 98 | 93 | 104 | ||||
Range (%) | 92-122 | 92-106 | 73-113 | 95-112 | |||||
1:4 | Average % of Expected | 107 | 107 | 97 | 112 | ||||
Range (%) | 100-115 | 100-116 | 79-117 | 104-118 | |||||
1:8 | Average % of Expected | 103 | 112 | 98 | 112 | ||||
Range (%) | 89-114 | 104-124 | 79-121 | 101-128 |
Assays for the Luminex platform are offered as High Performance Assays or Assays. The Luminex High Performance Assays are fully validated panels with a focused selection of analytes. They are available in We Mix, You Mix, and Predetermined formats. The Luminex Assays allows for the maximum number of analytes in a multiplex and is supplied as a premixed kit. View a table comparing the features and benefits of these bead-based multiplex assays.
Magnetic Luminex High Performance Assays are designed for use with the Luminex MAGPIX CCD Imager. Alternatively, kits can be used with the Luminex 100/200™ or Bio-Rad® Bio-Plex®, dual laser, flow-based sorting and detection platforms.
Analyte-specific antibodies are pre-coated onto color-coded microparticles. Microparticles, standards, and samples are pipetted into wells and the immobilized antibodies bind the analytes of interest. After washing away any unbound substances, a biotinylated antibody cocktail specific to the analytes of interest is added to each well. Following a wash to remove any unbound biotinylated antibody, Streptavidin-Phycoerythrin conjugate (Streptavidin-PE), which binds to the biotinylated detection antibodies, is added to each well. A final wash removes unbound Streptavidin-PE and the microparticles are resuspended in buffer and read using the Luminex or Bio-Plex analyzer. One laser is bead-specific and determines which analyte is being detected. A magnet in the analyzer captures and holds the superparaMagnetic microparticles in a monolayer. Two spectrally distinct Light Emitting Diodes (LEDs) illuminate the beads. One LED identifies the analyte that is being detected and the second LED determines the magnitude of the PE-derived signal, which is in direct proportion to the amount of analyte bound. Each well is imaged with a CCD camera. Kits can also be used with Luminex 100/200 or a Bio-Rad Bio-Plex dual laser, flow-based systems.
Scientific Data
Quantist software graph of signal intensity versus concentration for CD14 in the Human Biomarker A Luminex Panel
Background: CD14
CD14 is an acute phase glycoprotein that binds lipopolysaccharide endotoxins (LPS) from Gram-negative bacteria. It is primarily expressed on monocytes, macrophages, and neutrophils, and soluble forms can be secreted or shed from the cell surface. CD14 accepts LPS from its initial binding by LBP, and it can further transfer LPS to the TLR4/MD2 complex. LPS acctivation of TLR4 triggers inflammatory activation. CD14 also contributes to the response to TLR3 ligands, phosphoinositides, as well as the surfactant proteins SP-A and SP-D.
Product Datasheets
FAQs
No product specific FAQs exist for this product, however you may
View all Luminex Assay FAQsReviews for Human CD14 Magnetic Luminex® Performance Assay
There are currently no reviews for this product. Be the first to review Human CD14 Magnetic Luminex® Performance Assay and earn rewards!
Have you used Human CD14 Magnetic Luminex® Performance Assay?
Submit a review and receive an Amazon gift card.
$25/€18/£15/$25CAN/¥75 Yuan/¥2500 Yen for a review with an image
$10/€7/£6/$10 CAD/¥70 Yuan/¥1110 Yen for a review without an image